Potential wrong crosstalk measurement result with faulty vna cable

Дата публикации: 16-10-2026 12:36:42



Основное содержимое страницы с новостью.

Skip to main content
  • Thread starter Thread starter yefj
  • Start date Start date Friday, 7:36 AM
yefj
Messages
162
Reaction score
2
Hello , I have a PCB when I am doing NEXT tests of victim agressor on two differential lines.
The opposite end of the lines is terminated.

I am measuring sdd21 values around -40dB.
Very often I have problem with the cables.
suppose I have S11 S22 S33 S44 at -10dB which is good can I be comfident that the SDD21 result is accurate?
I had lots of situation where I have reflections inside the VNA cables which where shown by the impulse responce at t<0 of the sdd11.
So is it critical ? or good return loss at all ports means good cross talk?
Thanks.